OM 403PWR
6-DIGIT MULTIFUNCTION AC/DC V-A metr
The OM 403 series are 6-digit panel instruments designed for fast and easy setup, as well as accuracy and reliability.
Type OM 403PWR is a universal AC/DC V–A meter. The setup and operation of the device are user-friendly, despite its versatility, by incorporating two displays, touch keys with color navigation and haptic feedback, as well as our Setup Wizard integrated into the meter.
The device is based on a 32-bit processor and a multi-channel 24-bit ΔΣ ADC, which ensure high accuracy, stability and easy operation.
-
3-color LED main display, auxiliary display and bargraph
-
Voltage, Current, Active / Reactive / Apparent power
-
Input I-AC/DC: 60 / 150 / 300 mV, 1 /5 / 10 A
-
Input U-DC: 0…200 / 350 / 600 / 1000 V
-
Input U-AC: 0…150 / 250 / 450 / 700 V
-
Touch keys with haptic feedback and RGB backlighting
-
Teach-in, Digital filters, Tare, Mat. function, Linearization
-
DIN size 96 x 48 mm
-
Power supply 10...30 V AC/DC or 80…250 V AC/DC
Options
-
comparators (relays, open collectors)
-
data communication,
-
analog output
-
recording of measured values
Input
PWR - AC
Vstupní rozsah - I
0…0,25 A
<200 mV
Vstup 3
0…0,5 A
<200 mV
Vstup 3
0…1 A
<200 mV
Vstup 3
0…2,5 A
<200 mV
Vstup 4
0…5 A
<200 mV
Vstup 4
0…10 A
<200 mV
Vstup 4
0…60 mV
6,8 kOhm
Vstup 4
0…75 mV
6,8 kOhm
Vstup 4
0…100 mV
6,8 kOhm
Vstup 4
0…150 mV
6,8 kOhm
Vstup 4
0…300 mV
6,8 kOhm
Vstup 4
20…150 mV/kA (< 5000 A)
6,8 kOhm
Vstup 4
Vstupní rozsah - U
0…150 V
2,5 MOhm
Vstup 2
0…250 V
2,5 MOhm
Vstup 2
0…450 V
2,5 MOhm
Vstup 1
0…700 V
2,5 MOhm
Vstup 1
Vstupní kmitočet
<400 Hz
Měřené veličiny
| U | Napětí U [Vrms] |
| I | Proud I [Arms] |
| P | Činný výkon P [W] |
| S | Zdánlivý výkon S [VA] |
| Q | Jalový výkon Q [VAr] |
| cos φ | Účiník [-] |
| ° | Fázový posun φ [°] |
| f | Kmitočet f [Hz] |
| U_dc | DC složka - Napětí Udc [V] |
| I_dc | DC složka - Proud Idc [A] |
| E | Elektrická energie E [Wh] |
| Q | Elektrický náboj Q [Ah] |
PWR - DC
Vstupní rozsah - I
±0,25 A
< 25 mV
Vstup 3
±0,5 A
< 50 mV
Vstup 3
±1 A
<100 mV
Vstup 3
±2,5 A
<200 mV
Vstup 3
±5 A
<25 mV
Vstup 4
±10 A
<50 mV
Vstup 4
±60 mV
6,8 kOhm
Vstup 4
±75 mV
6,8 kOhm
Vstup 4
±100 mV
6,8 kOhm
Vstup 4
±150 mV
6,8 kOhm
Vstup 4
±300 mV
6,8 kOhm
Vstup 4
Vstupní rozsah - U
±200 V
2,5 MOhm
Vstup 2
±350 V
2,5 MOhm
Vstup 2
±600 V
2,5 MOhm
Vstup 1
±1000 V
2,5 MOhm
Vstup 1
Měřené veličiny
| U | Napětí U [V] |
| I | Proud I [A] |
| P | Výkon P [W] |
| E | Elektrická energie E [Wh] |
| Q | Elektrický náboj Q [Ah] |
Control input and output
No. of inputs
3, isolated, per contact, PNP/NPN, < 30 V
Input functions
| OFF | External input is disabled |
| TARE.~ | Activation of tare for channels A – C |
| CL.TR.~ | Tare reset for channels A – C |
| TR.+CL.~ | Tare activation + tare reset for channels A – C (> 1 s) |
| CUMUL.~ | Control of cumulative measurement for channels A – C |
| CL.M.M.~ | Min/Max value reset for channels A – C |
| CL.PK.~ | Peak value reset for channels A – C |
| SAMPLE | Takes a single measurement |
| HOLD | Stops the measurement process |
| HLD.MIN | Holds the Minimum value* |
| HLD.MAX | Holds the Maximum value* |
| HLD.M-M | Holds the MAX-MIN value* |
| HLD.AVG | Holds the Average value* |
| * | *The value is calculated from the period starting with the previous external input activation |
| LATCH | Releases Safety relay/Open Collector (IEC EN 61496) |
| RECORD | Records measured data |
| KEY.LCK | Locks (disables) the keys on the device |
Funkce vstupů - NÁHLED
| OFF | Bez funkce |
| HOD.~ | Zobrazení hodnoty kanálu A - C |
| MIN.~ | Zobrazení minimální hodnoty kanálu A - C |
| MAX.~ | Zobrazení maximální hodnoty kanálu A - C |
| PEAK.~ | Zobrazení špičkové hodnoty kanálu A - C |
| TARA.~ | Zobrazení Táry kanálu A - C |
| BRUTT.~ | Zobrazení BRUTTO kanálu A - C |
No. of output
1, isolated, open collector 30 V/100 mA
Output functions
Ready - Active when the device reports no error messages
Projection
Hlavní displej
Display
| Range | -99999…999999 |
| Type | three-color alphanumeric LED, 6-digit, 11-segment |
| Color | red / green / orange |
| display color can be fixed or variable according to two set limits (3 zones) | |
| Digit height | 14 mm |
| Decimal point | adjustable, floating or exponential display |
| Brightness | adjustable (10 / 25 / 50 / 75 / 100 %) |
| automatic adjustment according to ambient light |
Displayed values
| VAL.~ | Actual value of channels A - F |
| MIN.~ | Minimum value of channels A - F |
| MAX.~ | Maximum value of channels A - F |
| PEAK.~ | Peak value of channels A - F |
| ABS.~ | Absolute value of channels A - C |
Measurement units
| STAND. | The unit corresponding to the input type is displayed on the last two characters |
| NO.UNIT | No measurement units |
| USER | A user-defined unit is displayed on the last two characters |
Sekundární displej
Display
| Range | -99999…999999 |
| Type | single-color alphanumeric LED, 6-digit, 11-segment |
| Color | green |
| Digit height | 7 mm |
Displayed values
| OFF | Display is off |
| VAL.~ | Actual value of channels A - F |
| MIN.~ | Minimum value of channels A - F |
| MAX.~ | Maximum value of channels A - F |
| PEAK.~ | Peak value of channels A - F |
| ABS.~ | Absolute value of channels A - F |
| TARE.~ | Tare value of channels A - F |
| BRUTT.~ | Tare value + value channels A - F |
| CUMUL.~ | Cumulative sum of channels A - F |
| LIMIT | Display of the setpoint |
| ANALOG | Actual value of analog outputs |
| POPIS | User-defined description of the displayed channel |
| TIME | Current time |
| DATE | Current date |
Sloupcový displej
Display
| Range | 17 LEDs |
| Type | single-color LED, horizontal |
| Display mode | bar, trace, center (symmetric from center) |
| Color | orange |
Displayed values
| OFF | Display is off |
| VAL.~ | Actual value of Channels A - F |
| MIN.~ | Minimum value of channels A - F |
| MAX.~ | Maximum value of channels A - F |
| PEAK.~ | Peak value of channels A - F |
| ABS.~ | Absolute value of Channel A - C |
Info displej
Display
| Range | -9…99 |
| Type | single-color alphanumeric LED, 2-digit, 11-segment |
| Color | orange |
| Digit height | 7 mm |
Displayed values
| ERR | Error message |
Signalizační LED
Display
20 single-color LEDs indicating device functions and status (red, yellow, orange)
Ovládání
Tlačítka
přístroj se ovládá a nastavuje 5 dotykovými tlačítky
s RGB podsvícením a haptickou odezvou
tlačítkům "Left", "Up", "Down", "Enter" je možné přiřadit další volitelné funkce
s RGB podsvícením a haptickou odezvou
tlačítkům "Left", "Up", "Down", "Enter" je možné přiřadit další volitelné funkce
Menu
| PROFI | Plné nastavení přístroje s přístupem přes heslo |
| USER | Uživatelské menu libovolných položek vybraných z PROFI menu s přístupem bez hesla |
Volba funkce tlačítka
| OFF | tlačítko je bez funkce |
| TARUJ.~ | Aktivace Táry kanálu A - C |
| NUL.TR.~ | Nulování Táry kanálu A - C |
| NUL.M.M.~ | Nulování Min/Max hodnoty kanálu A - C |
| NUL.PK.~ | Nulování Peak hodnoty kanálu A - C |
| VZOREK | Spuštění jednorázového měření |
| ROZ.REL. | Rozepnuti relé/OC (Typ TRVALE) |
| ULOZ. H | Záznam naměřený hodnot - jednorázové uložení |
| PREP. D | Přepínání zobrazení kanálů na hlavním displeji |
Volba dočasného zobrazení
| OFF | tlačítko je bez funkce |
| HOD.~ | Hodnota kanálu A - C |
| MIN.~ | Minimální hodnota kanálu A - C |
| MAX.~ | Maximální hodnota kanálu A - C |
| PEAK.~ | Špičková hodnota kanálu A - C |
| TARA.~ | Zobrazení Táry kanálu A - C |
| BRUTT.~ | Hodnota Kanálu A - C + hodnota Táry A - C |
Volby rychlého přístupu na vybranou položku
| OFF | tlačítko je bez funkce |
| LIMIT.~ | Přímý přístup na položku LIMIT. 1 - 6 |
OM Link
Firemní komunikační rozhraní pro ovládání, nastavení a update firmware přístroje (BT, USB-C)
Instrument accuracy and functionality
Temperature coefficient
25 ppm/°C
Accuracy
±0,1 % z rozsahu
DC - U ≤350 V
±0,2 % z rozsahu
DC - U >350 V
±0,1 % z rozsahu
DC - I
0,1 % z rozsahu
AC - U ≤250 V
0,2 % z rozsahu
AC - U >250 V
0,1 % z rozsahu
AC - I
0,1 % z rozsahu
Kmitočet
0,4 % z rozsahu
P, S, Q, účiník, posuv
Rate
0,5/1,2/2,5/5 measur./s
Overload capacity
2x; 10x (t < 30 ms)
Linearization
lineární interpolací v 100 bodech (pouze přes OM Link)
Digital filters
| arithmetic mean | from 2…100 measurements |
| Plovoucí průměr | z 2…100 měření |
| exponential mean | from 2…100 measurements |
| rounding | setting projection step for display |
Functions
| Tare | designed to reset display on non-zero input signal |
| Offset | kompenzace rozdílu mezi měřenou a skutečnou/požadovanou hodnotou |
| Min/max value | registration of min./max. value achieved during measurement |
| Peak value | display shows only max. or min. values |
| Teach-in | semiautomatic regime of input calibration |
| Simulace | simulace vstupního signálu/funkce přístroje |
| Log | logování chyb a událostí |
| Odložený start | čas kdy se neprovádí měření po připojení přístroje k napájení |
Timer
časové a denní omezení provozu přístroje, funkcí a periferií
| Výběr | Záznam naměřený dat / Reléové výstupy / Analogový výstup / Dataový výstup / Externí vstupy |
| Časové okno | možnost nastavení až 3 nezávislých časových omezení |
| nastavuje se čas OD - DO a libovolný den v týdnu (Pondělí - Neděle) |
Math. functions
polynom, invertní polynom, 1/x, logarithm, exponential, power, root
Data record
Measured data recording into the memory
| Dlouhodový | čas-datum-hodnota displeje, < 100 000 údajů |
| Jednorázový | rychlý kontinuální záznam < 400 měř./s |
Time
přesnost < 1 minuta/rok
Watch-dog
reset after 500 ms
Calibration
at 25°C and 40 % r.h.
Comparators
Number of
2, 3, 4 nebo 6 výstupů
Type
Digital, menu adjustable, contact connect. < 30 ms
Volba zdroje signálu
| OFF | Relé je vypnuté |
| HOD.~ | Aktuální hodnota Kanálu A - C |
| MIN.~ | Minimální hodnota Kanálu A - C |
| MAX.~ | Maximální hodnota Kanálu A - C |
| PEAK.~ | Špičková hodnota Kanálu A - C |
| ABS.~ | Absolutní hodnota Kanálu A - C |
Mode
| AKT.NAD | Aktivní nad nastavenou hodnotou |
| Limita, Hystereze, Zpoždění sepnutí, Zpoždění rozepnutí | |
| AKT.POD | Aktivní pod nastavenou hodnotou |
| Limita, Hystereze, Zpoždění sepnutí, Zpoždění rozepnutí | |
| OKENKO | aktivní v nastaveném okně / pásmu |
| Začátek, Konec, Hystereze, Zpoždění sepnutí, Zpoždění rozepnutí | |
| DAVKA | aktivní v nastavených periodách |
| Peroda,Doba sepnutí, Zpoždění sepnutí |
Realy functions
| SW.ON | contact closed when active |
| SW.OFF | contact open when active |
| PULSE | closes once for a moment |
| LATCH | contact on untill disarmed |
Output
2 - 4x relé
přepínací kontakt (Form C)
250 VAC/30 VDC, 3 A
3 - 6x relé
spínací kontakt (Form A)
250 VAC/30 VDC, 3 A
3 - 6x otevřený kolektor
30 VDC/100 mA
Analogue outputs
Number of outputs
až 2 výstupy, izolované
Type
Isolated, programmable with a 16-bit D / A converter, type and range are menu selectable
Volby zdroje signálu
| HOD.~ | Aktuální hodnota Kanálu A - C |
| MIN.~ | Minimální hodnota Kanálu A - C |
| MAX.~ | Maximální hodnota Kanálu A - C |
| PEAK.~ | Špičková hodnota Kanálu A - C |
| ABS.~ | Absolutní hodnota Kanálu A - C |
Temperature coefficient
15 ppm/°C
Accuray
±0,02 % full scale
±0,03 % full scale (0…5 V)
±0,05 % full scale (0…2 V / 0…5 mA)
±0,03 % full scale (0…5 V)
±0,05 % full scale (0…2 V / 0…5 mA)
Rate
response to value change < 160 µs
Ranges
0…2 V
Err ~ 2,2 V
odporová zátěž > 1 kOhm
0…5 V
Err ~ 5,5 V
odporová zátěž > 1 kOhm
0…10 V
Err ~ 11 V
odporová zátěž > 1 kOhm
±10 V
Err ~ 11 V
odporová zátěž > 1 kOhm
0…5 mA
Err ~ 5,5 mA
kompenzace < 600 Ohm/12 V
0…20 mA
Err ~ 22 mA
kompenzace < 600 Ohm/12 V
4…20 mA
Err ~ 3,2 mA
kompenzace < 600 Ohm/12 V
Error state
Detection of current loop interruption
indikace chybového hlášení na všech rozsazích
indikace chybového hlášení na všech rozsazích
Data outputs
Number of outputs
1 outputs
Protocol
ASCII, MODBUS - RTU, MODBUS - TCP/IP, PROFINET
Data format
| Format | 8bit + parity + stop bit |
| Parity | none / even / odd |
| Stop bit | 1 / 1.5 / 2 |
Rate
600…921 600 Baud
10 Mbit/s, 100 Mbit/s
10 Mbit/s, 100 Mbit/s
Address
| ASCII | < 99 přístrojů |
| MODBUS | < 246 instruments |
Line termination
DIP SW - interní odpor 120 Ohm
Measured data logging
Výběr kanálů
Výběr kanálů pro záznam naměřených hodnot
| KANAL.~ | Volba záznamu pro kanálu A - C (Ano-Ne) |
Nastavení rozsahu záznamu naměřených hodnot
Tato funkce umožňuje omezit zaznamenávané hodnoty na základě
| VSE | Záznam všech hodnot |
| UVNITR | Záznam hodnot uvnitř nastaveného intervalu |
| VNE | Záznam hodnot mimo nastavenéný interval |
Způsob ukládání
Volba způsobu záznamu hodnot do paměti přístroje
| PERIODA | Časově řízené ukládání hodnot |
| Nastavitelná perioda v rozsahu 00:00:01…23:59:59 | |
| MAXI | Ukládání maximální rychlostí zápisu (< 400 záznamů/s) |
| SNIMEK | Jednorázové uložení aktuální hodnoty |
| TRIGG | Záznam hodnot s Pre-triggrem |
| Pro režim TRIGGER lze nastavit velikost paměti pro uložení dat přes spuštěcím impulzem |
Přepis naměřených záznamů
Volba způsobu ukládání hodnot do paměti přístroje
| NE | Přepis nejstarších uložených hodnot je vypnutý |
| ANO | Přepis nejstarších uložených hodnot je povolený |
Power supply
Power supply
10…30 V AC/DC, ±10 %, PF > 0,4, I < 40 A/1 ms, isolated
80…250 V AC/DC, ±10 %, PF > 0,4, I < 40 A/1 ms, isolated
80…250 V AC/DC, ±10 %, PF > 0,4, I < 40 A/1 ms, isolated
Current draw
< 9,4 W/9,2 VA
Power supply protection
Power supply is protected by a fuse inside the instrument
Mechanical properties
Dimensions
96 x 48 x 120 mm (w x h x d)
Panel cutout
90 x 45,5 mm (š x v)
Material
Noryl GFN2 SE1, non-inflammable UL 94 V-I
Operating conditions
Connection
connector terminal board, cable section < 1,5/2,5 mm2
Stabilization period
within 5 minutes after switch on
Operating temperature
-20°…60°C
Storage temperature
-20°…85°C
Operating humidity
< 95 % r.v., nekondenzující
Protection class
IP 65 (front panel only)
El. safety
EN 61010-1, A2
Dielectric strength
4 kVAC within 1 min. between pow. supply and input
4 kVAC within 1 min. between pow. supply and data/anal. output
4 kVAC within 1 min. between input and relay output
2,5 kVAC within 1 min.between input and data/anal. output
4 kVAC within 1 min. between pow. supply and data/anal. output
4 kVAC within 1 min. between input and relay output
2,5 kVAC within 1 min.between input and data/anal. output
Insulation resistance
for polution degree II, measuring cat. III.
| power supply | 670 V (ZI), 300 V (Di)* |
| input, output, PN | 300 V (ZI), 150 V (Di)* |
EMC
EN 61326-1
ČSN EN 62003 (Jaderná zařízení)
ČSN EN 62003 (Jaderná zařízení)
RoHS
IEC 63000 : 2018
Seismic capacity
IEC/IEEE 60980-344 ed. 1.0:2020, par 6,9
Mechanical durability
EN 60068-2-6 ed. 2:2008
Datasheets
| File name | File type | Version | Language | Download |
|---|---|---|---|---|
|
Datasheet OM403PWR
Katalogový list OM403PWR
|
PDF
|
2026.2 |
|
Download |
User manuals
| File name | File type | Version | Language | Download |
|---|---|---|---|---|
| Návod OM403PWR |
PDF
|
2026.01.05 | cs | Download |
Application notes
| File name | File type | Version | Language | Download |
|---|---|---|---|---|
|
Application note External inputs
Aplikační list Externí vstupy
|
PDF
|
2024.1 |
|
Download |
| Aplikační list Registry Modbus |
PDF
|
2026.001.0 | cs | Download |
| Application note OM403 |
PDF
|
2026.001.0 | en | Download |