OMB 300UNI
UNIVERSAL BARGRAPH - 30 LED
The OMB 200/300/500UNI model series are simple bargraphs designed for maximum efficiency and user comfort while maintaining their low cost.
The OMB 300UNI type is a multifunction instrument with the option of configuration for 5 different types of inputs easily configurable in the instrument menu.
The instrument is based on a single-chip microcontroller with A/D converter, which secures good accuracy, stability and easy operation of the instrument.
By selecting the mode of insertion of the front plexiglass (reverse/face) you may choose the required printing of scale for vertical or horizontal design of the instrument.
-
three-color bargraph - 30 LED
-
multifunction input (PM, OHM, RTD, DU)
-
digital filters, linearization
-
size of DIN 72 x 24 mm
-
power supply 10…30 VDC/24 VAC
Options
-
comparators
Input
PM
Range
menu selectable
0…20 mA
< 1,2 V
Input 1
4…20 mA
< 1,2 V
Input 1
0…2 V
182 kOhm
Input 2
0…5 V
182 kOhm
Input 2
0…10 V
182 kOhm
Input 2
OHM
Range
menu selectable
0…100 kOhm
0…100 kOhm
Connection
2-wire
RTD
Type
menu selectable
EU > 1 000 Ohm s 3 850 ppm/°C
-50°…450°C
Connection
2-wire
Ni
Type
menu selectable
Ni 1 000 s 5 000 ppm/°C
-50°…250°C
Connection
2-wire
DU
Range
Configuration is performed in two steps:
| 1. | setting numerical value for the beginning and end of the potentiometer range |
| 2. | slider position calibration for the beginning and end of the potentiometer runway |
Potentiometer power supply
2,5 VDC/6 mA, potentiometer resistance > 500 Ohm
External inputs
Number
1 input, on contact
Functions
| OFF | No function |
| HOLD | Stopping measurement (option: Display - Analog output - Limits) |
| LOCK | Blocking front panel keys |
Projection
Display range
30 LED
Display colour
red/green/orange
Signalizing LED
| 1, 2, 3 | Limits L1, L2, L3 (red) |
Display brightness
menu adjustable
Settings
Control
The device is controlled and set by five buttons located under the front panel
Selected functions can also be controlled via external inputs
Selected functions can also be controlled via external inputs
Menu
| LIGHT | contains only items necessary for basic device settings with password access |
| PROFI | Full device settings with password access (only via OM Link) |
OM Link
Company communication interface for operation, setting and update of device firmware
OM Link program is also designed for visualization and filing of meas. values from more instruments
OM Link program is also designed for visualization and filing of meas. values from more instruments
Instrument accuracy and functionality
Temperature coefficient
50 ppm/°C
Accuracy
±1 % of range + 1 digit
Rate
0,5/5/50/Max measur./s
Overload capacity
2x; 10x (t < 30 ms)
Wire compensation (RTD, OHM)
< 30 Ohm
Linearization
linear interpolation in 25 points (only via OM Link)
Digital filters
| exponential mean | from 2…100 measurements |
| rounding | setting projection step for display |
Watch-dog
reset after 25 ms
Calibration
at 25°C and 40 % r.h.
Comparators
Type
Digital, menu adjustable, contact connect. < 50 ms
Mode
| Hysteresis | switch. point, hysteresis band "Mez ± 1/2 Hys." and time (0 ... 99.9 s) determ. switching delay |
Output
1x bistable relay (250 VAC/30 VDC, 3A)
2x bistable relay (250 VAC/30 VDC, 3A)
3x bistable relay (250 VAC/30 VDC, 3A)
2x bistable relay (250 VAC/30 VDC, 3A)
3x bistable relay (250 VAC/30 VDC, 3A)
Power supply
Power supply
10…30 VDC/24 VAC, ±10 %, PF > 0,4, I < 45 A/1 ms, isolated
Current draw
< 2,3 W/2,4 VA
Mechanical properties
Dimensions
96 x 24 x 100 mm (w x h x d)
Panel cutout
90,5 x 21,5 mm (w x h)
Material
Noryl GFN2 SE1, non-inflammable UL 94 V-I
Operating conditions
Connection
connector terminal board, cable section < 1,5/2,5 mm2
Stabilization period
within 15 minutes after switch on
Operating temperature
-20°…60°C
Storage temperature
-20°…80°C
Protection class
IP40 (front panel only)
El. safety
EN 61010-1, A2
Dielectric strength
2,5 kVAC within 1 min. between pow. supply and input
4 kVAC within 1 min. between input and relay output
4 kVAC within 1 min. between input and relay output
Insulation resistance
for polution degree II, measuring cat. III.
| power supply | 300 V (ZI)* |
| input, output | 300 V (ZI), 150 V (Di)* |
EMC
EN 61326-1
Datasheets
| File name | File type | Version | Language | Download |
|---|---|---|---|---|
|
Datasheet OMB300UNI
Datasheet OMB300UNI
Datasheet OMB300UNI
Katalogový list OMB300UNI
|
PDF
|
2020.1 |
|
Download |
|
Datasheet OMB300UNI
Katalogový list OMB300UNI
|
PDF
|
2026.2 |
|
Download |
|
Datasheet OMB300UNI
Katalogový list OMB300UNI
|
PDF
|
2023.1 |
|
Download |
User manuals
| File name | File type | Version | Language | Download |
|---|---|---|---|---|
|
Operating instructions OMB300UNI
Návod OMB300UNI
Operating instructions OMB300UNI
|
PDF
|
2013.3.0 |
|
Download |
| Návod OMB300UNI |
PDF
|
2012.3.0 | cs | Download |
|
Operating instructions OMB300UNI
Návod OMB300UNI
Operating instructions OMB300UNI
|
PDF
|
2008.1.2 |
|
Download |
|
Operating instructions OMB300UNI
Návod OMB300UNI
|
PDF
|
2007.1.1a |
|
Download |
|
Operating instructions OMB300UNI
Návod OMB300UNI
|
PDF
|
2006.1.1 |
|
Download |
|
Operating instructions OMB300UNI
Návod OMB300UNI
|
PDF
|
2006.1.0 |
|
Download |
Certificates
| File name | File type | Version | Language | Download |
|---|---|---|---|---|
|
Certificate EU_Declaration_OMB2_3_500
Certifikát EU_Prohlaseni_OMB2_3_500
|
PDF
|
2022 |
|
Download |
|
Certificate Declaration_17050_OMB2_3_500
Certifikát Prohlaseni_17050_OMB2_3_500
|
PDF
|
2022 |
|
Download |
|
Certificate Seismic_test_OMB300
Certifikát Seizmicka_zkouzka_OMB300
|
PDF
|
2021 |
|
Download |